AcroPak™ 20 Filters & AcroPak 200 Capsule with Fluorodyne® II Membrane
Efficient Processing of Liquid
Volumes Typically Up To 20 L
- Fluorodyne II membrane is compatible with aqueous and many organic solvents
- AcroPak 20 filters are designed to add convenience to small- and medium-volume filtrations
- AcroPak 200 Fluorodyne II capsules provide fast processing of clean batches
- Double layer sterilizing membrane enhances reliability
- Upstream air vent prevents vapor lock
- Manufactured without the use of adhesives to minimize extractables
- Pre-sterilized by gamma irradiation
- Sterile filtration of media and buffers
- Point-of-use filtration for lab water
- Suitable for viscous, particulate solutions and serum containing media
- Applicable for many fluids containing dilute proteins or preservatives
- Scalable
Materials of Construction
- Filter Media: Fluorodyne II membrane (hydrophilic PVDF)
- Housing, Vent Plug: Polypropylene
- Sealing Technology: Thermal bonding
- Filling Bell: Polycarbonate
Pore Size
- 0.1 and 0.2 μm
Effective Filtration Area
- 20 and 200 cm2
Dimensions (Nominal)
- AcroPak 20 Filters
Housing Length: 8.3 cm (3.3 in.)
Housing Diameter: 6.7 cm (2.7 in.) - AcroPak 200 Capsule
Overall Length: 10.5 cm (4.1 in.)
Housing Diameter Without Vent: 5.3 cm (2.1 in.)
Housing Diameter With Vent: 6.7 cm (2.6 in.)
Inlet/Outlet Connections
- 6.4 - 12.7 mm (1/4 - 1/2 in.) diameter stepped hose barb with female luer slip interior and filling bell outlet
Maximum Operating Temperature and Pressure
- 60 °C (140 °F) at 1.0 bar (100 kPa, 15 psi)
- 4.1 bar (410 kPa, 60 psi) at 21 - 24 °C (70 - 75 °F)
Recommended Integrity Test Minimum Bubble Point – Water
- 0.2 μm Fluorodyne II membrane: 3.2 bar (320 kPa, 46 psi)
60% IPA/40% H2O (v:v)
- 0.1 μm Fluorodyne II membrane: 1.8 bar (180 kPa, 26 psi)
Typical Water Flow Rate mL/min at 0.1 bar (10 kPa) [mL/min/psi]
- AcroPak 20 Filters
0.1 μm Fluorodyne II Membrane: 13 (9)
0.2 μm Fluorodyne II Membrane: 26 (18) - AcroPak 200 Capsule
0.2 μm Fluorodyne II Membrane: 240 (170)
Typical Hold-Up Volume (with air purge)
- AcroPak 20 Filters < 2.5 mL
- AcroPak 200 Capsule < 6 mL
Bacterial Retention
- Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision
Endotoxin Level
- < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety
- Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization
- Sterilized by gamma irradiation and individually packaged
- AcroPak 20 filters can withstand one autoclave cycle at 131 °C (268 ºF) for 30 min
Contact Customer Service at 800-521-1520 for possible lead-time and ship date.
|