Acrodisc® Syringe Filters Optimized for Scale Up
Devices Constructed of the Same Materials
as Larger-Capacity Capsules and Cartridges
- Simplifies scale-up and minimizes revalidation; no need to change membrane materials during transition to pilot or production.
- Four membrane chemistries assure compatibility with a wide range of fluids:
- Supor® membrane has high flow rates and throughputs, and is ideal for solutions where low protein binding is required. Not recommended with some ketones.
- Fluorodyne® II membrane offers high flow rates and is ideal for applications where PVDF membrane is specified. Not recommended with some ethers.
- Ultipor® membrane provides broad solvent and chemical compatibility, and low extractables.
- Posidyne® membrane enhances bioburden and pyrogen removal from aqueous solutions.
- Integrity testable (water bubble point).
- Bacterial retention tested.
- Sterilized by gamma irradiation to eliminate potential contamination by EtO residuals.
- Drug development studies
- Determination of product compatibility and recovery
- Preliminary filterability testing
- Small-volume liquid sterilization
Materials of Construction
- Filter Media: Supor [hydrophilic polyethersulfone (PES)], Fluorodyne II (hydrophilic PVDF), Ultipor (Nylon 6,6), and Posidyne (positively-charged Nylon 6,6) membranes
- Housing: Polypropylene
Effective Filtration Area
- 2.8 cm2
Inlet/Outlet Connections
- Female luer lock inlet, male slip luer outlet
Typical Hold-Up Volume
- < 100 μL
Maximum Operating Temperature
- 60 °C (140 °F) at 2.1 bar (210 kPa, 30 psi)
Maximum Operating Pressure
- 5.4 bar (540 kPa, 80 psi) at ambient temperature
Typical Water Flow Rate
- mL/min at 2.1 bar (210 kPa, 30 psi)
- PN 4905: 130
- PN 4906: 78
- PN 4907: 130
- PN 4908: 77
Recommended Integrity Test Minimum Bubble Point - Water
- PN 4905: 3.5 bar (350 kPa, 51 psi)
- PN 4906, 4908, and 4907: 3.2 bar (320 kPa, 46 psi)
- PN 4902: 3.32 bar (332 kPa, 48 psi)
Bacterial Retention
- Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838-05, current revision
Endotoxin Level
- < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety
- Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization
- Sterilized by gamma irradiation and individually packaged
Contact Customer Service at 800-521-1520 for possible lead-time and ship date.
|
Your shopping cart is empty. |
| *US List Price applies only to products delivered within the US. These prices do not reflect duties, taxes or tariffs, which may apply to purchases made outside of the US. See our General Limited Warranty. |